Inmel 60
Calibrator, Multifunctional Meter - Calibrator For Automation Engineers
Parameters :-
1) Simulation And Measurement :
- Thermocouple Signals J, K, T, R, S, B, U, L, E, N
With Auto-Compensation Of Terminal Temperature.
- Signals In Thermocouple And Thermostat System.
- Resistance Thermometer Signals : Pt 100, Pt 500,
Pt 1000, Ni 100, Cu 100.
- Resistance Signals From 10 Ω
To 5 KΩ
- Voltages And Currents In Range : -11 V...
+11 V,
-22 mA...+22 mA.
2) Basic Error 0.025%.
3) STEP And RAMP Function.
4) Led Out 24 V Terminals.
5) RS232C Interface.
6) Available In The Option For Pressure Calibration.
Application :-
1) Checking And Calibration :
- Measuring Transducers Controlled By A Voltage Or
Current Signal, A Thermoelement Or Resistance Thermometer.
- Two Wire Transducers.
- Temperature Meters.
- Pressure Transducers.
- Manometers.
- Measuring And Control Lines.
Technical Parameters :-
Range
Resolution
Basic
Error (1)
Measurement
Generation
Voltage :
-88 mV to 88 mV
1 uV
±
(0.05% M + 4 d)
±
(0.05% S + 5 uV)
-220 mV to 220 mV
10 uV
±
(0.025% M + 3 d) ±
(0.025% S + 30 uV)
-2.2 V to 2.2 V
100 uV
±
(0.025% M + 2
d) ±
(0.025% S + 150
uV)
-11 V to 11 V (Measurement Up To 25V )
1 mV
±
(0.025% M + 2 d)
±
(0.025% S + 200 uV)
Current :
-22 mA to 22 mA
1 uA
±
(0.025% M + 3d)
±
(0.025% S + 3 uA)
Resistance :
10 Ω
to 1000 Ω
0.1 Ωͧ
±
(0.05% M + 2 d ),
±
(0.05% S + 30 mΩ),
(Measurement From 0 Ω)
Ip = 0.4 mA
Ip = 10 mA
10 Ω
to 5000 Ω
0.1 Ω
±
(0.05% M + 3 d ),
±
(0.05% S + 150 mΩ),
(Measurement From 0 Ω)
Ip = 0.4
mA
Ip = 1 mA
Resistance Thermometer :
-200 to 850 ºC
(Pt 100)
0.1 ºC
±
(0.025% M + 0.3 ºC)
±
(0.025% S + 0.1 ºC)
Ip = 0.4 mA
Ip = 10 mA
-200 to 850 ºC
(Pt 500, Pt 1000)
0.1 ºC
±
(0.025% M + 0.3 ºC)
±
(0.025% S + 0.1 ºC)
Ip = 0.4 mA
Ip = 1 mA
-50 to 180 ºC
(Cu 100)
0.1 ºC
±
(0.025% M + 0.3 ºC)
±
(0.025% S + 0.1 ºC)
Ip = 0.4 mA
Ip = 10 mA
-60 to 180 ºC
(Ni 100)
0.1 ºC
±
(0.025% M + 0.3 ºC)
±
(0.025% S + 0.1 ºC)
Ip = 0.4 mA
Ip = 10 mA
Thermocouple :
(J), -210 to 1200 ºC
(Fe-Cu Ni)
0.1º
C ±
(0.025% M + 0.5 ºC)
±
(0.025% S + 0.5 ºC)
(K), -270 to 1370 ºC
(Ni Cr-Ni Al)
0.1 ºC
±
(0.025% M + 0.5 ºC)
±
(0.025% S + 0.5 ºC)
(T), -270 to 400 ºC
(Cu-Cu Ni )
0.1 ºC
±
(0.025% M + 0.5 ºC)
±
(0.025% S + 0.5 ºC)
(R), -50 to 350 ºC
(Pt Rh 13-Pt)
0.1 ºC
±
(0.05% M + 1.5 ºC)
±
(0.05% S + 1.5 ºC)
350 to 1769 ºC
(Pt Rh 13-Pt )
0.1 ºC
±
(0.05% M
+ 1 ºC)
±
(0.05% S +
1.0 ºC)
(S), -50 to 0 ºC
(Pt Rh 10-Pt )
0.1 ºC
±
(0.05% M + 1.5 ºC)
±
(0.05% S +
1.5 ºC)
0 to 1769 ºC
(Pt Rh 10-Pt )
0.1 ºC
±
(0.05% M + 1 ºC)
±
(0.05% S + 1
ºC)
(B), 200 to 1820 ºC
(Pt Rh 30-Pt Rh6)
0.1 ºC
±
(0.025% M + 1 ºC)
±
(0.025% S + 1 ºC)
(U), -200 to 600 ºC
(Cu-Cu Ni )
0.1 ºC
±
(0.025% M + 0.5 ºC)
±
(0.025% S + 0.5 ºC)
(L), -200 to 900 ºC
(Fe-Cu Ni )
0.1 ºC
±
(0.025% M + 0.5 ºC)
±
(0.025% S + 0.5 ºC)
(E), -50 to 700 ºC
(Ni Cr-Cu Ni )
0.1 ºC
±
(0.025% M + 0.5 ºC)
±
(0.025% S + 0.5 ºC)
(N), -270 to 1300 ºC
(Ni Cr Si Ni Si)
0.1 ºC
±
(0.025% M + 0.5 ºC)
±
(0.025% S + 0.5 ºC)
Pressure (2) :
-100 to 150 kPa
±
0.1%
M
±
0.1% S
0 to 7 MPa
±
0.1% M
±
0.1% S
Note : Ip = Measuring Current
M = Measurement Value
S = Setting Value
(1) = During 12 Months After
Calibration, At Temperature Of 23 ±
5 ºC
(2) = Customer Establishes The
Measuring Range Of The Pressure Transducer
General Technical Data :-
1) Power Supply : 230V ±
10%, 50 Hz, 15 VA
2) Battery / Power Consumption : 12V / 1.9 Ah / 4 VA Max.
3) Dimensions : 230 x 90 x 205 mm
4) Weight : 2.2 Kg
5) Operation Time : Min. 4 Hours
6) Battery Charging Time : 12 Hours
7) Accessories : Power Supply, Protective Carrier Bag, Measuring Wires
8) Working Temperature : +5... +40 ºC
9) Relative Humidity : 20...80%