Inmel 20 / 21 Calibrator

Parameters :
1) Simulation Of Pt 100 Sensor.
2) Simulation Of J, K, S * Thermocouple With Terminal Temperature Auto-Compensation.
3) Simulation Of Systems : Thermocouple And Thermostat.
4) Voltage Generation From -0.5V Up To 11 V.
5) Current Generation From -2mA Up To 22 mA.
6) Basic Error 0.1%
7) RS232C Interface.

Applications :
1) Checking And Calibration :
     - Measuring Transducers Controlled By A Voltage Or A Current Signal, A Thermocouple Or A Resistance Thermometer.
     - Temperature Meters.
     - Measuring And Control Lines.

Note : Custom-made Calibrator Version With Optionally Chosen Types Of Thermocouples Can Be Arrange.
Technical Parameters :

    Range                 Setting             Resolution        Basic Error (1)      Max. Load             Add. Error           Add. Error
                               Range                                     ± (% S + % R)                                At Max. Load          At 10 ºC
                                                                                                                                         
± (% R )          ± (% S + % R)

  Pt 100 (2)       -200...+850
ºC         1 ºC             (0.08 + 0.02)           15 mA                      -                   (0.08 + 0.02)
     5 V (2)          -0.5...+5.5 V         0.001 V         (0.08 + 0.02)            22 mA                  (0.01)             (0.02 + 0.005)
      10 V              -1...+11 V             0.01 V          (0.08 + 0.02)           22 mA                   (0.01)            (0.02 + 0.005)
      5 mA          -0.5...+5.5 mA       0.001 mA        (0.08 + 0.02)      -5.5...+11 V              (0.005)           (0.02 + 0.005)
     20 mA           -2...+22 mA          0.01 mA         (0.08 + 0.02)      -5.5...+11 V              (0.005)           (0.02 + 0.005)
 J (Fe-CuNi)     -210...+1200
ºC        1 ºC         (0.08+0.02+1 ºC)  output R = 3.5Ω     (I load x 3.5Ω)     (0.03+0.01)
K(NiCr-NiAl)  -270...+1372
ºC         1 ºC         (0.08+0.02+1 ºC)  output R = 3.5Ω     (I load x 3.5Ω)     (0.03+0.01)
 S(PtRh10-Pt)   -50...+1769
ºC          1 ºC         (0.08+0.02+1 ºC)  output R = 3.5Ω     (I load x 3.5Ω)     (0.03+0.01)

Note : S - Set Value
          R - End Value Of the Range ( for Pt 100 : 850
ºC )
        (1) - Basic Error During 12 Month After The Calibration At Temperature Of 23
ºC ±ºC
        (2) - 5 V Range Is Available In Inmel 20 Calibrator : Pt 100 Range Is Available In Inmel 21 Calibrator.

Calibrator Version :
1) Inmel 20 :
    - With The Range 5 V Without Pt 100.
    - With The Range 5 V Without Pt 100 With Interface RS232C, Without Accumulator.

2) Inmel 21 :
    - With The Range Pt 100 Without 5V.
    - With The Range Pt 100 Without 5 V With Interface RS232C, Without Accumulator.

General Technical Data :-
1) Power Supply : 230V ± 10%, 50 Hz, 15 VA
2) Battery / Power Consumption : 12V / 1.9 Ah / 4 VA Max.
3) Dimensions : 230 x 55 x 165 mm
4) Weight : 2.4 Kg
5) Operation Time : Min. 4 Hours
6) Battery Charging Time : 12 Hours
7) Accessories : Power Supply, Protective Carrier Bag, Battery
8) Working Temperature : +5... +40 ºC
9) Relative Humidity : 20...80%