Prova-Inmel 123

Parameters :
1) Voltages Generation : 0 - 100.00 mV, 0 - 1.000 V, 0 - 12.000 V.
2) Current Generation : 4 - 20 mA, 0 - 20 mA, 0 - 24 mA.
3) Simulation Of J, K, S, T Thermocouples.
4) Frequency Generation : 1 - 62500 Hz.
5) Voltage Basic Error : 0.08%.
6) Current Basic Error : 0.4%.
7) "STEP" And "RAMP" Functions.
8) Percentage Settings (mA, mV, V).
9) Sound Signalling Of Overload (mA, mV, V).

Application :
1) Checking And Calibration :
     - Measuring Transducers Controlled By A Voltage Signal, A Current Signal Or A Thermocouple.
     - Temperature Meters.
     - Measuring And Control Lines.
     - Impulse Counters.

Technical Parameters :

1) Current Ranges ( Load <= 1 k , 24 V Loop Supply ).

                 Range                                 Resolution                       Basic Error

4-20 mA, 0-20 mA, 0-24 mA                  1 uA                      ± 0.025% ± 3 uA


Beeper Warning When Outputs Is Open And Specified Current Output > 1 mA.

2) Voltage Ranges ( 1 mA Supply Current ).

                 Range                                 Resolution                       Basic Error

             0 - 100 mV                               10 uV                       ± 0.05% ± 30 uV
               0 - 1 V                                   100 uV                     ± 0.05% ± 300 uV
              0 - 10 V                                    1 mV                       ± 0.05% ± 3 mV

Beeper Warning When Outputs Is Short And Specified Voltage Output > 10 mV.

3) Thermocouple Ranges ( Resolution 1 ºC, 1 ºF, 5 Min After Switching On, Load >= 1 kΩ )

                     Range                                 Basic Error                         Range                            Basic Error

K : -200...0 ºC     (Ni Cr-Ni Al)                ± 1.1 ºC                    K :  -328...32 ºF                     ± 2.0 ºF
K :  0...1370 ºC   (Ni Cr-Ni Al)                 ± 0.8 ºC                   K :    32...2400 ºF                   ± 1.5 ºF
J  : -100...0 ºC         (Fe-Cu Ni)                ± 0.9 ºC                    J :    -148...32 ºF                     ± 1.6 ºF
J  :   0...760 ºC         (Fe-Cu Ni)                ± 0.7 ºC                    J :     32...1400 ºF                   ± 1.2 ºF
S :   0...1760 ºC    (Pt Rh 10-Pt)                ± 2.0 ºC                    S :    32...3200 ºF                   ± 3.0 ºF
T : -200...0 ºC          (Cu-Cu Ni)               ± 1.0 ºC                    T :  -328...32 ºF                      ± 1.8 ºF
T :   0...400 ºC         (Cu-Cu Ni)                ± 0.8 ºC                    T :    32...752 ºF                     ± 1.5 ºF

4) Frequency Ranges ( Load >= 1 k
Ω ).

                      Range                                  Resolution                      Basic Error

                   1 - 125 Hz                                 1 Hz                           
± 0.04 Hz
              126 - 62500 Hz                              1 Hz                    
± 0.01 % ± 0.04 Hz

Electrical Specification ( 23 
ºC ± 5 ºC, 5 Minutes After The Power Is On )

General Technical Data :-
1) Battery Type : 9V Alkaline Battery.
2) Power Consumption : 60 mA, 180 mA ( depends on load ).
3) Dimensions : 88 x 160 x 26 mm
4) Weight : 330 g
5) Display : 4 and 5 digits.
6) Storage Temperature : -20 to 60 ºC.
7) Accessories : Carrying case, User manual, K type thermocouple connector, External Battery Pack Holder 1.5V AA,
                          9V Battery, Test leads with alligator clip.
8) Working Temperature : 0... 50 ºC
9) Working  Humidity : less than 85%